WARNING: Ensure that surgical gloves and protective eyewear are worn whenever
cleaning or maintaining any part of the suction system.


Step 1:

Fill the Pulse Cleaner with 2.5 litres of warm water (NB: each marking point on the inside of the Pulse Cleaner marks 1 litre, so you should fill it to half way between the 2nd and 3rd lines).

IMPORTANT - always fill the Pulse Cleaner with water first, DO NOT put in the Pulijet PLUS first as this will create strong odours.

Please download a flash player to see the illustration.

Step 2:

Place the Pulijet PLUS bottle on a flat and steady surface and remove the lid.

Please download a flash player to see the illustration.

Step 3:

Squeeze the bottle of Pulijet PLUS until the fluid reaches the top of the cap.

Please download a flash player to see the illustration.

Step 4:

Remove pressure from the bottle. The Pulijet PLUS should fall to the marking on the clear lid.

Please download a flash player to see the illustration.

Step 5:

Pour the Pulijet PLUS fluid contained in the top into the Pulse Cleaner.

Please download a flash player to see the illustration.

Step 6:

Connect the large and small suction tubes to the Pulse Cleaner and wait until the solution has been evacuated (there will be a small quantity that remains in the Pulse Cleaner).

Please download a flash player to see the illustration.

Step 7:

Pour the remaining fluid in the Pulse Cleaner into the spittoon.